IBP Monitoring
15-7
IBP alarm limits
Pressure Label
Max. Upper (mmHg)
Min. Lower (mmHg)
Step (mmHg)
ART 300
0
1
PA 120 -6 1
CVP 40 -10 1
RAP 40 -10 1
LAP 40 -10 1
ICP 40 -10 1
P1 300 -50 1
P2 300 -50 1
SCALE
ADJUST >>
This option allows you to open the following menu and set
the positions of the high, reference and low scales for the
two waveforms.
Figure 15-7 IBP Scale Adjust Menu
1. HI: Determines the high scale. The reference scale changes when the high scale
is changed.
2. LO: Determines the low scale. The reference scale changes when the low scale
is changed.
3. VAL: Determines the reference scale and position. The high and low scales
remain unchanged when the reference scale is changed.
NOTE
z
The high scale should not be smaller than the low scale; the low scale
should not be larger than the high scale.
z
The waveform amplitude changes when the high and low scales are
adjusted.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...