NIBP Monitoring
13-10
NOTE
z
The calibration of the NIBP measurement should be performed every two
years or performed according to the Hospital Procedure.
13.4.2 Testing for Air Leakage
The PNEUMATIC option is used to test air leakage. When the NIBP cuff is
connected, select this option to start the NIBP inflation and test whether the air
leakage occurs in the airway. If the test is passed, no prompt information will be
displayed; if the test is not passed, corresponding prompt information will be
displayed in the NIBP parameter window.
To test air leakage, use the following procedure:
1. Connect the NIBP cuff with the NIBP cuff connector of the monitor.
2. Wrap the cuff around a cylinder of a proper size, as shown below.
Figure 13-4 NIBP Leakage Test
3. Select the PNEUMATIC option, and the information “Pneum testing…” is
displayed at the lower left corner of the NIBP parameter window.
4. After approximately 20 seconds, the monitor will automatically open the
deflate valve, ending the test.
5. If no information appears on the bottom of the NIBP parameter area, it
indicates the airway is in good condition and an air leak does not exist.
However if the information “PNEUMATIC LEAK” appears, it indicates the
airway may have an air leak. In this case, check for loose connections. After
confirming all connections are secure, perform the test again.
If there is still a failure, contact our Customer Service.
Monitor
NIBP cuff
connector
Hose
Cylinder
Cuff
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...