SpO
2
Monitoring
12-17
A blood pressure cuff is inflated on the same extremity as the one with a SpO
2
sensor attached;
The patient has hypotension, severe vasoconstriction, severe anemia, or
hypothermia;
There is arterial occlusion proximal to the sensor;
The patient is in cardiac arrest or is in shock.
12.3.5 SpO
2
Setup Menu
Selecting the SPO
2
label in the parameter window opens the following menu.
Figure 12-7 MASIMO SpO
2
Setup Menu
You can perform the following settings in this menu.
ALM
SpO
2
alarm on/off status
ON: When a SpO
2
alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a SpO
2
alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right
of the SPO
2
label.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...