ECG/RESP Monitoring
11-12
11.3 ECG Setup Menu
Select the ECG label in the parameter windows. The following menu appears.
Figure 11-7 ECG SETUP Menu
In this menu, you can perform the following settings:
HR ALM
Heart rate alarm on/off status
ON: When a heart rate alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a heart rate alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right of
the ECG label.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
ALM REC
Alarm recording
ON: When a heart rate alarm occurs, the monitor enables the
recording;
OFF: When a heart rate alarm occurs, the monitor does not
enable the recording.
ALM HI
Upper alarm limit
Determines the upper ECG alarm limit.
ALM LO
Lower alarm limit
Determines the lower ECG alarm limit.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...