System Structure and Assembly/Disassembly
4-29
4.3.13 Replacement of the travel switch (with the cable)
a)
Remove the main panel assembly and minor panel assembly (see section 4.3.4
and section 4.3.5), and remove the fixing screws of the cast-aluminum cover to
take out the cast-aluminum cover.
b)
Remove two screws (M2X10 cross panhead screw) fixing the travel switch (with
cable), replace it with new cable of the travel switch, and connect the cable
properly.
c)
According to the reversed order, install the travel switch properly.
Lock head
Travel
switch
Screw
Screws fixing lock head