System Structure and Assembly/Disassembly
4-21
(3) Loosen the fixing screws of the encoder PCBA and main panel PCBA, and
remove the encoder PCBA and main panel PCBA, and replace it with a new
one.
(4) Remove silicone keys from the main panel PCBA:
a)
The silicone keys are installed in the main panel PCBA through the
locating poles of the keys and the corresponding locating holes in the panel
PCB, and they can be pulled up from the panel by hand.
b)
When the locating poles are pulled up for installation of a new panel, note
that each locating poles shall be pulled properly.
Encoder PCBA
Fixing screws
Main
panel
PCBA
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...
Страница 191: ...P N 2105 20 40473 V10 0...