Checks
6-2
version.
6.1.2.5
Check the system operation
Check that the menus and windows can pop up normally.
Check that measurements can be performed in each exam mode, such as OB measurements
and cardiac measurements, etc.
Check that exam modes can be normally switched.
Check that display modes can be normally switched.
Check that the transducers can be normally switched.
6.2
Function checks
6.2.1 Flowchart for function checks
Check the displayed time
and date
↓
Check the initial operations
↓
Check general operations
↓
Check keys on the control
panel
↓
Check the trackball
↓
Check the peripherals of the
main unit
↓
Check the fan
6.2.2 Content of the function checks
1)
Checking displayed time and date
Check that the displayed time and date are correct.
If not, correct them.
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...
Страница 191: ...P N 2105 20 40473 V10 0...