System Structure and Assembly/Disassembly
4-49
3)
Remove 2 M3X8 countersunk screws for each transducer mounting bracket
and shielding plate, remove 2 M2.5X8 panhead screws for the transducer
socket and transducer mounting bracket, and remove the transducer mounting
brackets (6 in total).
4)
Remove 12 M3x8 screws for the transducer board and shielding plates, and
separate the shielding plates and transducer board.
4.3.20 Replace the inlet dust net
a)
Loosen two nondetachable screws for two inlet dust nets and main unit housing,
and take out the inlet dust net from the left side of the system.
M3X8
countersunk
screw
M2.5X8 panhead
screw
Transducer
mounting
bracket
Transducer board
Shielding plate
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...
Страница 191: ...P N 2105 20 40473 V10 0...