System Structure and Assembly/Disassembly
4-25
c) Loosen the screw (M4X12) on the keyboard connector clip, and remove the cable
joint of the keyboard to take out the keyboard.
d) Remove the fixing screws (4 M4X8 panhead screws with washers) of remaining
slide tracks and keyboard support, and install the slide tracks to the new keyboard
assembly.
e) Insert the connector of the keyboard cable to the new keyboard assembly, and
tighten the screw on the keyboard connector clip.
f) Hold the keyboard by hands, align the tracks, push the keyboard evenly to the end,
and repeatedly push and pull to check if the keyboard is properly installed.
4.3.10 Replacement of the light control PCBA
a) Remove the fixing screws of the main panel assembly (see 4.3.4), lift the front of
the main panel assembly, and remove the socket of the light control PCBA on the main
panel assembly.
Screw on the
keyboard
connector clip
Slide track
Slide track
Fixing screw of
slide track
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...
Страница 191: ...P N 2105 20 40473 V10 0...