System Structure and Assembly/Disassembly
4-52
bottom of the right side), draw out the transformer from the left side, disconnect
the input and output cables of the transformer, and take out the transformer.
4.3.22 Remove the fan assembly
a)
Remove the inlet fan of the power main unit box.
1)
Remove the main unit rear cover.
2)
Remove 2 M3
×
8 screws of the inlet fan assembly.
3)
Hold the grips, and draw out the inlet fan assembly.
b)
Remove the outlet fan of the power main unit box.
4 M5x10 inner hexagonal screws
Cables
M3X8 screw
Inlet fan
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...