Adjust the LCD monitor after changing or replacing the main board:
a)
Follow step 1, 2 and 3 in section 7.2.3.1 to enter the System Maintenance Menu
and select the Monitor Test item.
b)
Press the left direction key on the minor keyboard to display the following screen.
Press <1> and <4> buttons and hold for at least 3s until the blue menu displays on
the screen, select “autojust” via button <1> and <2>, and press <4> to confirm; the image
Tilt the monitor to the position as shown in the
figure, and hold it by two sides and lift it in the
direction of F, until it is disconnected.