7-1
7
System Maintenance
7.1
Cleaning
7.1.1 Flowchart for cleaning
Unplug the power plug from the wall outlet
↓
Remove the accessories
↓
Remove the housing of the main unit
↓
Clean interior of the system and fan
↓
Clean covers and control panels
↓
Clean the monitor
↓
Clean peripherals and accessories
↓
Clean the power cable and other cables
7.1.2 Content of the cleaning
1)
Interior of the system and fan
Unplug the power plug from the wall outlet.
Remove the accessories: transducer, transducer holder and printer, etc.
Remove the housing of the main unit and shielding covers and IO box, thus the internal
PCBAs, power module and monitor assembly can be seen. If the PCBAs are normal, do not
remove them.
Use a big plastic bag to cover the main unit.
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...
Страница 191: ...P N 2105 20 40473 V10 0...