System Structure and Assembly/Disassembly
4-47
4.3.19 Remove the transducer board
a)
Remove the transducer assembly: remove 22 M3x8 screws fixing the transducer
assembly and main unit, hold the two grips to draw out the board; care shall be
taken to avoid damage of interior connectors.
b)
Remove the shielding cover of the transducer board and aluminum shielding
frames of the transducer sockets
1)
Remove 19 M3x8 screws fixing the shielding cover of the transducer board
and aluminum shielding frames, remove 8 M3x8 screws fixing the transducer
board and shielding cover, and then remove the shielding cover of the
transducer board.
22 M3x8
screws
Grips
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...