Checks
6-7
item
check content
procedure
tool
Mechanical
safety
4) Remove the rear cover of the
control panel and neck cover
of the monitor, check by a
wrench to confirm that the
fixing screws are free of
looseness; check by eyes to
confirm that the cables are
free of scratch.
Philips
screwdrive
r,
inner
hexagonal
wrench
Fixation
of
the
peripherals
Use the wrench to check and
confirm that the peripherals
are free of looseness.
Cable connection
1) Check by eyes to confirm that
the cables, inside or outside
the system, are free of
damage and scratch.
Philips
screwdrive
r
2) Check by hand to confirm that
the cables inside the system
are free of looseness and
falling off.
Appearance
of
the
transducers
Check by eyes to confirm that the
transducers are free of crack,
peeling, looseness or
damage.
Other mechanisms
Check to confirm that other
mechanisms are free of crack
or exposure of conductive
parts.
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...
Страница 191: ...P N 2105 20 40473 V10 0...