System Structure and Assembly/Disassembly
4-46
e)
Remove the M3X8 screw, disconnect the cables, and remove the ECG base
assembly and ECG mounting cover.
f)
Remove M3X8 screws, and replace the ECG board or footswitch connecting
PCBA.
M3X8 screw
ECG base assembly
ECG mounting cover
M3X8 screw
ECG base
ECG
board
or
footswitch
connecting
board