Checks
6-9
6.4.2.4
Recorded images
The printed images shall be archived as per the above description.
6.4.3 Final operation checks and image archive
6.4.3.1
Image operation check
Check the image operations using each transducer in each mode.
6.4.3.2
Image printing and archiving
The printed images shall be archived with the recorded data.
6.4.3.3
Flowchart for checks
Check functions
↓
Check image operations
↓
Image archiving
Perform the above checks for each transducer.
6.4.3.4
Check content
1)
Checking functions
Check the mode change, measurements and relevant information.
2)
Checking image operations
Perform image operations using each transducer in each mode.
3)
Image archiving
Print the images and archive the images.
6.4.4 Checking color flow images
The service engineers shall be trained by Mindray’s professionals, and the method is to use a
transducer to examine and capture a color carotid flow image. The color flow image shall be
evaluated depending on whether the flow is full, whether there is an artifact or overflow, and
whether the edge is smooth.
Содержание DC-6
Страница 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Страница 2: ......
Страница 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Страница 23: ...System Overview 2 4 2 2 LCD Monitor...
Страница 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Страница 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Страница 97: ...System Structure and Assembly Disassembly 4 32 10 screw caps 13 M4 8 screws Rear cover assembly...
Страница 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Страница 138: ...Checks 6 4 6 3 Safety Checks 6 3 1 Electric safety Refer to Appendix D Electrical Safety Inspection...
Страница 187: ...D 13 ELECTRICAL SAFETY INSPECTION 7 Mains on Applied Part Leakage LIMITS For BF ECG Input and transducer 5000 A...
Страница 191: ...P N 2105 20 40473 V10 0...