System Structure and Assembly/Disassembly
4-14
4.3.4 Disassembly of the main panel
a)
Remove the three holders on the right side of the main panel.
b)
Remove three M3x8 screws on the side of the main panel assembly and five
M4x12 screws at the bottom (The figure below doesn’t show the main unit but
shows the positions of the screws.)
1. Adjusting nut of the spring
1
Holder