System Maintenance
7-4
7.2
Software Maintenance
Note: If power is off during upgrading process, the upgrading operation will fail and the system
cannot be reset.
7.2.1 Back up the preset data
Prior to software maintenance, confirm that all relevant data must be those used in this
system; otherwise it may result in system damage. At the same time, the user shall back up
the preset data as per the procedures below.
1.
Press the F10 key on the keyboard to enter the preset menu, as shown in the figure
below:
2.
Move the cursor onto “Preset Data”, and select “Save”, as shown in the figure
below: