Lake Shore Model 321 Autotuning Temperature Controller User’s Manual
Remote Operation
4-19
Curve Commands
(Continued)
Example:
00, STANDARD DRC-D ,N,31,
0.00000,499.9,0.19083,365.0,
0.24739,345.0,0.36397,305.0,
0.42019,285.0,0.47403,265.0,
0.53960,240.0,0.59455,220.0,
0.73582,170.0,0.54606,130.0,
0.95327,090.0,1.00460,070.0,
1.04070,055.0,1.07460,040.0,
1.09020,034.0,1.09700,032.0,
1.10580,030.0,1.11160,029.0,
1.11900,028.0,1.13080,027.0,
1.14860,026.0,1.07200,025.0,
1.25070,023.0,1.35050,021.0,
1.63590,017.0,1.76100,015.0,
1.90660,013.0,2.11720,009.0,
2.53660,003.0,2.59840,001.4,
6.55360,000.0[term]
ECUR11
Edit or Add A Data Point In User Curve 11.
Input:
ECUR11,X.XXXXX,YYY.Y
Returned:
Nothing
Remarks:
Fill in the point to be added or edited, where X.XXXXX is voltage and YYY.Y is
the temperature in kelvin. If the Model 321 does not recognize either the units
value or the temperature value, it will assume that you are entering a new point
and place it in the proper ascending position.
Example:
If the point to be edited was input as 0.19083,364.0 and should have been
0.19083,365.0, input the command as follows:
ECUR 11,0.19083,365.0[term]
The Model 321 will recognize the units field and replace that data point with the
new temperature value.
KCUR11
Delete User Curve 11 Data Command.
Input:
KCUR11
Returned:
Nothing
Remarks:
Will delete all data stored for the User Curve 11.
SCAL
SoftCal™ Voltage Entry.
Input:
SCAL 12,X.XXXXX,Y.YYYYY,Z.ZZZZZ
Returned:
Nothing
Remarks:
Stores the SoftCal™ voltage values at 4.2 K, 77.32 K, and 300 K,
where X.XXXX = 4.2 K voltage, Y.YYYYY = 77.32 K voltage, and
Z.ZZZZZ = 300 K voltage.
Example:
SCAL 12,1.6260,1.0205,0.5189[term]