Lake Shore Model 321 Autotuning Temperature Controller User’s Manual
4-16
Remote Operation
4.2.3 Curve
Commands
This paragraph provides a detailed description of each Curve Command. The commands allow the
user to verify existing curves added at the factory or enter and delete the user defined curve over the
interface.
ACUR
Assign Curve Number.
Input:
ACUR XX
Returned:
Nothing
Remarks:
Fill in curve parameter with an integer from 0 through 12. Refer to Table 2-2 for a
listing of sensor curve numbers.
ACUR?
Curve
Number
Query.
Input:
ACUR?
Returned:
XX
Remarks:
Value returned is an integer from 00 through 12, corresponding to the currently
selected sensor curve number. Refer to Table 2-2 for a listing of sensor curve
numbers.
ACOMP
Set Room Temperature Compensation
(Model 321-04 Thermocouple Only).
Input:
ACOMP 0
or
ACOMP 1
Returned:
Nothing
Remarks:
Select temperature compensation parameter with 0 for off or 1 for on.
ACOMP?
Room Temperature Compensation Query
(Model 321-04 Thermocouple Only).
Input:
ACOMP?
Returned:
0
or
1
Remarks:
Returns current room temperature compensation status where 0 = off and 1 = on
ATYPE?
Input Type Query.
Input:
ATYPE?
Returned:
SI
,
PT
,
TC
, or
ER
Remarks:
Returns input type where SI = silicon diode, PT = platinum, TC = thermocouple,
and ER = error (improper switch setting).