Lake Shore Model 321 Autotuning Temperature Controller User’s Manual
Service
6-11
6.8.2.4
Analog Output Calibration
Use the following procedure to perform the analog output calibration.
1. Attach test leads to TP3 (gnd1) and TP4 (Analog out).
2. Press and hold the Analog Out key until the Model 321 enters the Analog output calibration
mode.
3. Press the Up arrow key until Zero is selected.
4. With the DMM in DC voltage mode, adjust R22 until the voltage is 0.000 ±0.001 VDC.
5. Press the Up arrow key until Span is selected.
6. Adjust R18 until the voltage is 10.000 ±0.001 VDC.
7. Press Escape on the Model 321 to exit the Analog output calibration mode.
8. Set power switch to off (
O
). Disconnect power cord and test resistor.
9. Replace top of enclosure and secure with four screws.
10. Replace back bezel and use 5/64 hex key to tighten two screws attaching bottom panel to unit.
6.9 MODEL 321-02 (PLATINUM RESISTOR) CALIBRATION
There are two Model 321-02 (Platinum Resistor) calibrations provided. If you have a 100
Ω
, 0.01%
25 ppm precision resistor available, use the procedure in Paragraph 6.9.1. If no precision resistor is
available, use the optional procedure in Paragraph 6.9.2.
6.9.1 Model 321-02 Calibration (With Precision Resistor)
The Model 321-02 calibration with precision resistor is presented in four parts. First is list of test
equipment in Paragraph 6.9.1.1. Second is the test setup in Paragraph 6.9.1.2. Third is the input
calibration in Paragraph 6.9.1.3. Finally, analog output calibration is provided in Paragraph 6.9.1.4.
These procedures will be performed with power supplied to the unit. Due to the presence of lethal
currents, only personnel experienced in working with live circuits should perform this procedure.
6.9.1.1 Test
Equipment
This procedure requires the following test equipment:
•
Digital Multimeter (DMM) with a 4½-digit display, capable of 4-lead resistive measurements.
•
Precision Resistor, 100
Ω
, 0.01%, 25 ppm/°C.
6.9.1.2 Test
Setup
Calibration test setup is as follows:
WARNING:
Lethal line voltage is present inside the box. Only qualified personnel should attempt
to perform this procedure.
1. Set power switch to off and disconnect power cord from rear of unit.
2. Use 5/64 hex key to remove four screws attaching top panel to unit.
3. Use 5/64 hex key to loosen two screws attaching bottom panel to unit.
4. Carefully remove back bezel.
5. Slide top panel back and remove from unit.