Lake Shore Model 321 Autotuning Temperature Controller User’s Manual
Service
6-5
6.5 OPERATING SOFTWARE EPROM REPLACEMENT
The operating software for the Model 321 is contained on two Erasable Programmable Read Only
Memory (EPROM) Integrated Circuits (ICs). The EPROM are numbered U22 (Master) and U19 (Slave).
The EPROMs will also have a label on top identifying the software version and date. See Figure 6-10.
Please use the following procedure to replace either one or both the operating software EPROMs.
WARNING:
To avoid potentially lethal shocks, turn off the instrument and disconnect it
from the AC power line before performing this procedure. Only qualified
personnel should perform this procedure.
1. Set power switch to off and disconnect power cord from rear of unit.
2. Use 5/64 hex key to remove four screws attaching top panel to unit.
3. Use 5/64 hex key to loosen two screws attaching bottom panel to unit.
4. Carefully remove back bezel.
5. Slide top panel back and remove from unit.
6. Locate software EPROM U22 (Master) or U19 (Slave) on main circuit board. Note orientation of
existing EPROMs (circular notch on front of IC). See Figure 6-10.
7. Use IC puller to remove existing EPROM(s) from socket.
8. Noting orientation of new EPROM(s), use IC insertion tool to place new EPROM(s) into socket.
9. Replace top of enclosure and secure with four screws.
10. Replace back bezel and use 5/64 hex key to tighten two screws attaching bottom panel to unit.
11. Reconnect power cord to rear of unit and set power switch to on.
12. Proceed to Paragraph 2.12 and perform the initial setup and system checkout. When replacing the
operating software EPROM(s), all operating parameters will be returned to the factory default
settings.
U19
Front
Rear
Fuse
Power Inlet
Operating Software
EPROMs
U22
Transformer
Master
Master
Slave
Slave
C-321-6-9.eps
Figure 6-9. Location Of Operating Software EPROMs