TOS9200/9201
Specifications 8-11
*6
The DCW (DC withstanding voltage testing) function is provided only for TOS9201.
Item
Speci
fi
cations
Indicator
240x64-dot LCD. Displays settings, measurements and judgement results.
Testing
function
ACW
Executes AC withstanding voltage test
DCW (*6)
Executes DC withstanding voltage test
IR
Executes insulation resistance test
AUTO (*6)
Automatically executes AC withstanding voltage test (ACW), DC withstand-
ing voltage test (DCW), insulation resistance test (IR), and earth continuity
test (EC). The TOS6200 (Earth Continuity Tester) is necessary for earth con-
tinuity testing.
Memory
function
ACW
Maximum of 100
DCW (*6)
Maximum of 100
IR
Maximum of 100
AUTO
Maximum of 100 test patterns, with each having up to 100 steps (total of 500
steps)
Backup battery life
At least 3 years (at 25
˚
C)
MEASURE MODE
Displaying the measured current in withstanding voltage testing and the
measured insulation resistance in insulation resistance testing can be
selected as shown below
NORM
Displays, during the test, the measured current in withstanding voltage test-
ing and the measured insulation resistance in insulation resistance testing
MAX/MIN
Displays, during the test, the maximum measured current in withstanding
voltage testing and the minimum measured insulation resistance in insula-
tion resistance testing during one test
PASS HOLD TIME
The time during which a PASS judgement is held can be set from 0.2 s to
10.0 s at a resolution of 0.1 s, or to HOLD.
TEST TIME
MOMENARY
Executes a test only while the START switch is pressed
FAIL MODE
Cancels to reset the FAIL or PROTECTION status produced by the STOP
signal with remote control
DOUBLE ACTION
Starts a test only when the START switch is pressed within approximately
0.5 s after the STOP switch is pressed
KEYLOCK
Shifts to the status in which no keys other than the START and STOP
switches are accepted
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......