TOS9200/9201
Precautions on Handling 2-5
2.5
Dangerous States of Failed Tester
Typical possible dangerous states of the tester are as shown below and in which
cases the most dangerous situation that
“the high test voltage remains delivered
and won't be turned off!”
may occur. When this situation has occurred, immedi-
ately turn OFF the power switch and disconnect the AC power cable from the AC
line receptacle.
• The DANGER lamp does not go out despite you have pressed the
STOP switch.
• The DANGER lamp does not light up despite the pointer of the ana-
log voltmeter is de
fl
ected indicating that the output voltage is being
delivered.
Also when the tester is in other malfunctioning states than the above, there is a pos-
sibility that the output voltage is delivered irrespective of your proper operating pro-
cedure. Never use the tester when it has failed.
•
Keep the tester away of other people until you call our service engineer for
help.
• Immediately call Kikusui distributor/agent. It is hazardous for an unquali-
fi
ed person to attempt to troubleshoot any tester problem.
2.6
To Ensure Long-Term Use Without Failures
The withstanding voltage-generating block of the tester is designed to release half
the rated amount of heat, in consideration of the size, weight, cost, and other factors
of the tester. The tester must therefore be used within the ranges speci
fi
ed below. If
you deviate from these ranges, the output block may be heated to excess, activating
the internal protection circuit. Should this happen, wait until the temperature
returns to the normal level.
Output requirements for withstanding voltage testing
(Output time = voltage rise time + test time + voltage fall time)
Ambient
temperature
Upper current
Pause
Output time
t
≤
40
˚
C
AC
50< i
≤
110 mA At least as long as the output time
Maximum of 30 minutes
i
≤
50 mA
Not necessary
Continuous output possible
DC
5< i
≤
11 mA
At least as long as the output time
Maximum of 1 minute
i
≤
5 mA
At least as long as the judgement
wait time (WAIT TIME)
Continuous output possible
WARNING
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......