3-56 Basic Operations
TOS9200/9201
3.10.1 SYSTEM 1
Setting the measurement mode (MEAS MODE)
A leakage-current measurement mode can be selected.
NORM
: Ordinary measurement mode
MIN/MAX : The measurement mode for holding the maximum current in
withstanding voltage testing and the minimum resistance
recorded after the WAIT TIME has elapsed in insulation resis-
tance testing
1.
Using the F1 key (MEAS) or the keys, move the cursor to
MEAS MODE.
2.
Using the rotary knob, select NORM or MIN/MAX.
Setting the pass hold time (PASS HOLD)
The hold time can be set to 0.2 s through 10.0 s (at a resolution of 0.1 s) for a PASS
judgement.
When HOLD is set, a PASS judgement is held until the STOP switch is pressed.
1.
Using the F2 key (PASS) or the keys, move the cursor to
PASS HOLD.
2.
Using the rotary knob, set the pass hold time.
Turning the momentary function ON/OFF (MOMENTARY)
• Releasing the START switch with MOMENTARY on is equivalent to pressing the
STOP switch. As a result, in a programmed automatic test, the test is unable to
proceed to the next step if the step interval time is set to HOLD.
When MOMENTARY is on, the test continues as long as the START switch is held
down.
When MOMENTARY is set to ON, the tester performs a test only for as long as the
START switch remains pressed. Since this operation keeps the operator’s hands
con
fi
ned to the front panel of the tester or the optional START switch during a test,
MEAS
PASS
SYSTEM 1
SYSTEM 1
MEAS MODE
: NORM
PASS HOLD
: 10.0s
MOMENTARY
FAIL MODE
DOUBLE ACTION
F1 F2 F3 F4 F5
OFF ON
NOTE
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......