3-48 Basic Operations
TOS9200/9201
Example of a DC withstanding voltage test (DISP2)
Example of an insulation resistance test (DISP1)
Example of an insulation resistance test (DISP2)
• In insulation resistance testing, when the resistance value to exceed 99.9 G
Ω
is
detected, “99.9”
fl
ashes on the LCD.
• In insulation resistance testing with the lower judgment function off, even if set-
tings are made so that the test voltage divided by the lower resistance exceeds 1.1
mA, testing can be performed. However, when the voltage value which is out of
the range of ±(10 % of s50 V) is detected during the test, the measurement
voltage value
fl
ashes on the LCD.
DISP
TEST
1.00
kV
1.4
s
12
A
NORM
DCW
DCW
999 M
Measured current
DISP
50
V
10.0
G
3.0
s
IR
IR
LOWER
OFF
OFFSET
OFF
TIMER
OFF
TEST
NORM
Measured resistance
DISP
TEST
50
V
1.4
s
12.0
G
NORM
IR
IR
10 nA
Measured resistance