3-50 Basic Operations
TOS9200/9201
Fig.3-19 Timing chart for PASS judgement
(The start voltage is set on the DCW.)
Fig.3-20 Timing chart for PASS judgement (IR)
FAIL judgement
Example of an AC withstanding voltage test (FAIL)
■
A
C/DC withstanding v
olta
ge
testing
If the lo
wer judgement function is of
f, the test ends with a F
AIL judgement when a current
lar
ger than the upper current is detected.
After a fe
w milliseconds, the po
wer is then cut of
f.
100 ms
Rise Time
Test Time
In PASS judgment
START
HV ON
TEST
PASS
READY
Output
voltage
Start
voltage
Rise Time
Test Time
In PASS judgment
START
HV ON
TEST
PASS
READY
Output
voltage
DISP
2.50
kV
1.0
mA
3.0
s
ACW
ACW
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
NORM
60Hz
FAIL