TOS9200/9201
Basic Operations
3-51
If the lo
wer judgement function is on, the test ends with a F
AIL judgement when the
tester detects a current lar
ger than the upper current or a current smaller than the
lo
wer current.
After a fe
w milliseconds, the po
wer is then cut of
f.
When lo
wer and upper judgements are made, “
↓
F
AIL” and “
↑
F
AIL” appear
,
respecti
v
ely
, at the top right of the LCD.
At the same time, the F
AIL LED lights on the indicator and a b
uzzer sounds.
T
o
cancel the F
AIL judgement, press the ST
OP switch (the F
AIL judgement is output
until the ST
OP switch is pressed).
The measurement results are displayed until the ST
OP switch is pressed.
The F
AIL judgement other than that described abo
v
e is a CONT
A
CT F
AIL (“
→←
F
AIL”) judgement that is made when the contact f
ailure is detected while a scanner
is used.
•
In
A
C withstanding v
oltage testing, a lo
wer judgement is not made during a v
olt
-
age rise time (RISE
TIME) or a v
oltage f
all time (F
ALL
TIME).
•
In DC withstanding v
oltage testing, no lo
wer judgement is made during a v
oltage
rise time (RISE
TIME).
•
In DC withstanding v
oltage testing, an upper judgement is not made until the
W
AIT
TIME has elapsed to eliminate the ef
fects of a char
ge current in capaciti
v
e
DUTs.
■
Insulation resistance testing
If the upper judgement function is on, the test ends with a F
AIL judgement when a
resistance lar
ger than the upper resistance is detected.
After a fe
w milliseconds, the
po
wer is then cut of
f.
If the lo
wer judgement function is on, the test ends with a F
AIL judgement when the
tester detects a resistance smaller than the lo
wer resistance.
After a fe
w millisec
-
onds, the po
wer is then cut of
f.
Using a lo
wer and upper judgement, “
↓
F
AIL” and “
↑
F
AIL” appear
, respecti
v
ely
,
at the top right of the LCD.
At the same time, the F
AIL LED lights up on the indica
-
tor and a b
uzzer sounds.
T
o cancel the F
AIL judgement, press the ST
OP switch (the F
AIL judgement is out
-
put until the ST
OP switch is pressed).
The measurement results are displayed until the ST
OP switch is pressed.
The F
AIL judgement other than that described abo
v
e is a CONT
A
CT F
AIL (“
→←
F
AIL”) judgement that is made when the contact f
ailure is detected while a scanner
is used.
•
In insulation resistance testing, an upper judgement is not made during a v
oltage
rise time (RISE
TIME).
•
In insulation resistance testing, a lo
wer judgement is not made until the
W
AIT
TIME has elapsed to eliminate the ef
fects of a char
ge current in capaciti
v
e DUTs.
NOTE
NOTE
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......