2-2
Precautions on Handling
TOS9200/9201
2.2
Action When in Emergency
In case of an emer
genc
y (such as electric shock hazard or b
urning of DUT), tak
e the
follo
wing actions.
Y
ou may do either (a) or (b)
fi
rst. But be sure to do both.
(a)
T
urn OFF the po
wer s
witc
h of the tester
.
(b)
Disconnect the A
C po
wer cor
d of the tester fr
om the A
C line recep
-
tac
le
.
2.3
Precautions on Testing
■
W
earing Insulation Glo
ves
When handling the tester
, be sure to wear
insulation glo
v
es in order to protect your
-
self ag
ainst high v
oltages. If no insulation glo
v
es are a
v
ailable on your mark
et,
please order Kikusui distrib
utor/agent for them.
■
Precautions f
or P
ausing T
ests
When changing test conditions, press the ST
OP switch once to tak
e precautions. If
you are not going to resume the test soon or if you are lea
ving the
T
est area, be sure
to turn-OFF the PO
WER switch.
Fig.2-1
Suspending testing and oper
ation
■
Items Char
ge
d Up to Dang
er
ous High
V
olta
ge
s
When in test, the DUT
, test leadwires, probes, and output terminals and their vicini
-
ties can be char
ged up to dangerous high v
oltages. Ne
v
er touch them when in test.
•
The vin
yl sheaths of the alligator clips of the test leadwires which are sup
-
plied accompan
ying the tester ha
v
e no suf
fi
cient insulation f
or the high test
v
oltages
.
Ne
v
er touch them when in test.
POWER switch
STOP switch
WARNING
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......