4-12 Using Terminals and Connectors
TOS9200/9201
INTERLOCK-signal input requirements
Open: Terminal-to-terminal current of 0.5 mA or less
Short-circuited: Terminal-to-terminal current of 1 mA or more
Connecting the interlock jumper
1.
Insert a screwdriver into A, and open B.
2.
Insert the interlock jumper into B. Take care to prevent the jumper coat-
ing from becoming caught in B.
3.
Lightly pull on the jumper to con
fi
rm that it is securely connected.
4.
Take the same steps for the positive (+) and negative (-) terminals.
Then, short-circuit (+) and (-).
To connect a safety device, use the following cable and switch types:
Compatible cables:
• Solid wire: ø0.65 (AWG22)
• Stranded wire: 0.32 mm
2
(AWG22); element-wire diameter: ø0.18 or more
• Removed coating: 10 mm
Rated switch voltage and current:
• DC 30 V, 0.1 A or more
Circuit for reference
When the door is opened, its contacts open, thus setting the INTERLOCK signal
high. This activates the interlock function.
Connection to the door open/close switch
A
B
Shank diameter: ø3; tip width: 2.6 mm
10 mm
Door open/close switch