8-4 Specifications
TOS9200/9201
Item
Speci
fi
cations
TOS9200
TOS9201
Judgement function
Judgement method/action
Judge-
ment
Judgement method
Display
Buzzer SIGNAL I/O
UPPER
FAIL
When the tester detects a current exceeding
the upper current, it cuts off the output and
makes an UPPER FAIL judgement. In DC
withstanding voltage testing, however, no
judgement is made until the judgement wait
time (WAIT TIME) has elapsed.
The FAIL
LED lights
up.
Displayed on
the LCD
ON
Outputs the U
FAIL signal
LOWER
FAIL
When the tester detects a current below the
lower current, it cuts off the output and
makes a LOWER FAIL judgement. How-
ever, no judgement is made during the volt-
age rise time (RISE TIME) or voltage fall
time (FALL TIME) in AC withstanding
voltage testing.
The FAIL
LED lights
up.
Displayed on
the LCD
ON
Outputs the L
FAIL signal
PASS
When the preset time has elapsed without
any abnormalities, the tester cuts off the out-
put and makes a PASS judgement.
The PASS
LED lights
up.
Displayed on
the LCD
ON
Outputs the
PASS signal
• The PASS signal is output at the timing preset on PASS HOLD. If HOLD is set, the PASS signal
is output continuously until the STOP signal is input.
• The UPPER FAIL signal and the LOWER FAIL signal are output continuously until the STOP
signal is input.
• The FAIL and PASS buzzer volumes are adjustable. However, they cannot be adjusted individu-
ally, as they are set in common.
Setting range for the upper
current (UPPER)
0.01 mA to 110 mA AC
0.01 mA to 110 mA AC /
0.01 mA to 11 mA DC
Setting range for the lower
current (LOWER)
0.01 mA to 110 mA AC
(With the LOWER OFF function)
0.01 mA to 110 mA AC /0.01 mA to 11 mA
DC (With the LOWER OFF function)
Judgement accuracy (*3)
±(3% of s 20
μ
A)
[After the offset cancel function is activated, if the scanner is mounted]
Current detection method
The absolute current values are integrated and compared with the reference value.
Response switching func-
tion
The current detection response for UPPER FAIL judgement can be set to FAST/
MID/SLOW (for AC withstanding voltage testing only).
Time
Setting range for the volt-
age rise time (RISE TIME)
0.1 s to 200 s
Setting range for the volt-
age fall time (FALL TIME)
0 s to 200 s (Valid only with PASS
judgement)
0 s to 200 s (Valid only with PASS judge-
ment in AC withstanding voltage testing)
Setting range for the test
time (TEST TIME)
0.3 s to 999 s With the TIMER OFF function
Accuracy
± (100 ppm + 20 ms)
Setting range for the judge-
ment wait time (WAIT
TIME)
––––––––
0.3 s to 10 s
(Only for DC withstanding voltage testing)
[(RISE TIME + TEST TIME) > WAIT
TIME]
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......