TOS9200/9201
Basic Operations 3-69
3.13.3 Suspending the program
To suspend the test while running a program, press the STOP switch.
To start the program from the beginning, press the START switch.
3.13.4 Judgement on the program
■
P
ASS judg
ement
When the interv
al time for each step is set to other than HOLD, the program mak
es
a P
ASS judgement if a P
ASS judgement is made for e
v
ery step (only if the pro
-
grammed repeat test is set to END).
A P
ASS judgement is made after the interv
al time for the pre
vious step has elapsed.
The tester then returns to the READ
Y status.
If the interv
al time for an
y step is set to HOLD, the test does not proceed to the ne
xt
step unless the ST
AR
T switch is pressed.
■
F
AIL judg
ement
Once a F
AIL judgement is made while the program is being run, the test stops at
that step.
Check the f
ailed step and press the ST
OP switch.
If the ST
AR
T switch is pressed ag
ain, the program starts from the be
ginning.
3.13.5
Exiting the program
T
o e
xit the PR
OGRAM mode in order to return to the normal mode, press the
A
CW
k
e
y (or the DCW or IR k
e
y).
The LED on the
A
UT
O k
e
y goes of
f, and the test-condition setting screen appears.
3.14
Key Lock
The k
e
y lock function is used to pre
v
ent the unintentional alteration of test condi
-
tions.
T
o lock the panel settings, press the SHIFT + LOCAL/KEY LOCK k
e
ys.
On the panel, only the ST
AR
T and ST
OP switches are operable.
While the k
e
y lock function is on, “KEY LOCK” is displayed on the LCD.
T
o cancel the k
e
y lock function, press the SHIFT + LOCAL/KEY LOCK k
e
ys.
1.00
kV
10.0
mA
3.0
s
ACW1
ACW1
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
OFF
READY
UPPER
60Hz
KEY
LOCK
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......