TOS9200/9201
Maintenance
7-3
7.5
Troubleshooting
Not all problems in
v
olv
e mechanical f
ailures. Before requesting repair service,
recheck your problem.
Find your problem in the list belo
w
, and correct it as instructed. If it still persists or
if you cannot
fi
nd it in the list, contact Kikusui distrib
utor/agent
Prob
lem
Chec
kpoint
Items f
or ref
erence
P
age
The tester does not start when
the PO
WER is turned on.
•
Is the po
wer cord disconnected?
•
Is the fuse brok
en?
The LCD screen does not appear
when the PO
WER is turned on.
•
Is the contrast le
v
el too lo
w?
•
Is the ambient temperature too lo
w?
“General Items of Speci
fi
cations”
A panel k
e
y does not w
ork.
•
Is the K
e
y Lock function w
orking?
•
Is the panel k
e
y remotely controlled
from the REMO
TE terminal or SIG
-
N
AL I/O?
•
Is the panel k
e
y remotely controlled
from outside via the GPIB or RS-
232C?
The ST
ART switch does not
work.
• Is the STOP signal input?
"Chapter 4 Using Terminals and Con-
nectors"
• Is the tester in the PROTECTION,
PASS, or FAIL status?
• Are you currently making settings
for the system and interface or edit-
ing a program?
"3.10 System Settings"
"3.11 Interface Settings"
"3.13 Program"
• Are you currently storing data in or
recalling it from the panel memory?
• Is the double-action function valid?
Turning the double-action function
ON/OFF (DOUBLE ACTION)
• Is “READY” displayed?
• Is the ENABLE signal for SIGNAL
I/O at the low level?
The output voltage does not
reach 2.6 kV.
• Is the output voltage range set to
AUTO?
Setting the output-voltage ranges (V
RANGE)
The output voltage does not
reach the preset level.
• Is RISE TIME set to too large a
value?
RISE TIME settings for each test
The fan does not work.
• The revolutions per minute are con-
trolled based on the internal temper-
ature. The fan stops revolving if the
temperature is too low.
If “OVERHEAT” appears without
the fan revolving, a fault has
occurred.
The high-voltage scanner does
not work.
• Is the interface cable connected?
The TOS6200 does not start up. • Is a cross cable being used as the
interface cable?
• Are the RS-232C settings correct?
If they are incorrect, correct them.
Then, reset the POWER switch.
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...