7-2
Maintenance
TOS9200/9201
7.3
Maintenance
•
T
o replace components
, the tester co
v
er m
ust be opened, b
ut only b
y our
ser
vice representativ
e
.
Contact y
our Kikusui agent/distr
ib
utor if it is neces
-
sar
y to perf
or
m replacement w
or
k.
■
High-v
olta
ge
rela
y
The high-v
oltage relays inside the tester are consumable.
Although service life
depends on the usage en
vironment, we recommend that the high-v
oltage relays be
replaced e
v
ery one million tests.
W
e also recommend re
gular internal inspections
and cleaning at the same interv
als.
■
Cooling fan
The service life of the cooling f
an is approximately 60,000 hours.
The re
v
olutions per minute of the cooling f
an are controlled based on the internal
temperature, and its service life depends partly on the conditions under which it is
used. Replace the cooling f
an after approximately 60,000 hours. In addition, con
-
duct internal inspection and cleaning of the tester at the same time.
■
Bac
kup batter
y
The tester uses a lithium battery for memory backup.
When the battery’
s po
wer supply runs out, such data as test conditions cannot be
stored in memory
. Depending on the conditions under which it is used, replace the
backup battery e
v
ery three years. In addition, conduct internal inspection and clean
-
ing of the tester at the same time.
7.4
Calibration
T
o maintain the accurac
y of measuring instruments for an e
xtended period, calibra
-
tion should be performed re
gularly
.
•
The tester gener
ates v
oltages as high as A
C 5 kV and DC 6 kV
.
Calibr
ation
is e
xtremely dangerous;
contact y
our Kikusui agent if calibr
ation ser
vices
are required.
WARNING
WARNING
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......