TOS9200/9201
Using Terminals and Connectors 4-13
4.4
STATUS OUT Connector
The STATUS OUT connector is an output connector to be connected to the warning
light.
Choose warning statuses on the System screen, from among H.V ON, TEST, PASS,
READY, U FAIL, L FAIL, CONTACT FAIL, PROTECTION, and POWER ON.
If two or more statuses are chosen at a time, the logical sum of them, is outputted.
For settings on the System screen, see "3.10 System Settings".
Maximum output voltage: DC 24 V; maximum output current: 100 mA
Compatible cables:
• Solid wire, ø0.65 (AWG22)
• Stranded wire, 0.32 mm
2
(AWG22); element wire diameter: ø0.18 or more
Connecting to the connector
1.
Remove approximately 10 mm of coating from the tip of the signal
cable.
2.
Insert a screwdriver into A, and open B.
3.
Insert the cable into B. Take care to prevent the coating from becoming
caught in B.
4.
Gently pull on the cable to con
fi
rm that it is securely connected.
•
Do not extract a current exceeding the maximum rated current of 100 mA
from the +24 V internal power supply. Once a current exceeds the maxi-
mum rated current, the overcurrent protection function activates and “24V
OCP”
fl
ashes on the LCD.
The tester enters the PROTECTION status. If the overcurrent protection
function activates, turn off the POWER switch to prevent an overcurrent.
If an overcurrent
fl
ows for an extended period, the internal circuit may mal-
function.
A
B
Shank diameter: ø3; tip width: 2.6 mm
10 mm
CAUTION
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......