218
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
13
Acquisition Control
Oscilloscope Bandwidth and Sample Rate
An oscilloscope's bandwidth is typically described as the lowest frequency at
which input signal sine waves are attenuated by 3 dB (-30% amplitude error).
At the oscilloscope bandwidth, sampling theory says the required sample rate is f
S
= 2f
BW
. However, the theory assumes there are no frequency components above
f
MAX
(f
BW
in this case) and it requires a system with an ideal brick-wall frequency
response.
However, digital signals have frequency components above the fundamental
frequency (square waves are made up of sine waves at the fundamental frequency
and an infinite number of odd harmonics), and typically, for 500 MHz bandwidths
and below, oscilloscopes have a Gaussian frequency response.
Figure 35
Theoretical Brick-Wall Frequency Response
f
S
f
N
-3dB
Attenuation
Frequency
0dB
Содержание DSOX3012T
Страница 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Страница 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Страница 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Страница 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Страница 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Страница 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Страница 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Страница 294: ...294 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 17 Digital Voltmeter and Counter...
Страница 300: ...300 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 18 Frequency Response Analysis...
Страница 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Страница 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Страница 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Страница 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...
Страница 416: ...416 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 25 CAN LIN Triggering and Serial Decode...
Страница 426: ...426 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 26 CXPI Triggering and Serial Decode...
Страница 436: ...436 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 27 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 30 Manchester NRZ Triggering and Serial Decode...
Страница 510: ...510 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 32 SENT Triggering and Serial Decode...
Страница 520: ...520 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 33 UART RS232 Triggering and Serial Decode...