206
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
11
Triggers
When two non-overlapping zones are used, their conditions are ANDed to become
the final qualifying condition.
When two overlapping zones have the must intersect condition, the zones are
ORed. When two overlapping zones have different conditions, Zone 1 takes
precedence and Zone 2 is not used. In this case, Zone 2 will have no fill (that is,
neither solid nor shaded) to indicate that it is not being used.
The zone qualified trigger feature is incompatible with, and will disable, the XY and
Roll horizontal time modes and the Averaging acquisition mode.
NOTE
Keep in mind that the TRIG OUT signal comes from the oscilloscope's hardware trigger. The
TRIG OUT signal indicates when there is a trigger (acquisition) that is evaluated for zone
intersection, not when an acquisition meets the zone qualification and is plotted on the
oscilloscope's display.
Содержание DSOX3012T
Страница 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Страница 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Страница 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Страница 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Страница 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Страница 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Страница 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Страница 294: ...294 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 17 Digital Voltmeter and Counter...
Страница 300: ...300 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 18 Frequency Response Analysis...
Страница 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Страница 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Страница 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Страница 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...
Страница 416: ...416 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 25 CAN LIN Triggering and Serial Decode...
Страница 426: ...426 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 26 CXPI Triggering and Serial Decode...
Страница 436: ...436 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 27 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 30 Manchester NRZ Triggering and Serial Decode...
Страница 510: ...510 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 32 SENT Triggering and Serial Decode...
Страница 520: ...520 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 33 UART RS232 Triggering and Serial Decode...