Vertical Controls
3
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
81
To adjust the vertical scale
1
Turn the large knob above the channel key marked
to set the vertical
scale (volts/division) for the channel.
You can also do this using the touchscreen. See
The vertical scale knob changes the analog channel scale in a 1-2-5 step
sequence (with a 1:1 probe attached) unless fine adjustment is enabled (see
change the vertical scale knob's coarse/fine adjustment setting"
The analog channel Volts/Div value is displayed in the status line.
The default mode for expanding the signal when you turn the volts/division knob
is vertical expansion about the ground level of the channel; however, you can
change this to expand about the center of the display. See
To adjust the vertical position
1
Turn the small vertical position knob ( ) to move the channel's waveform up or
down on the display.
You can also make this adjustment using the touchscreen. See
Waveform Offsets By Dragging Ground Reference Icons"
on page 53 and
Channels On/Off and Open Scale/Offset Dialogs"
The offset voltage value represents the voltage difference between the vertical
center of the display and the ground level (
) icon. It also represents the voltage
at the vertical center of the display if vertical expansion is set to expand about
ground (see
“To choose "expand about" center or ground"
To specify channel coupling
Coupling changes the channel's input coupling to either
AC
(alternating current)
or
DC
(direct current).
Содержание DSOX3012T
Страница 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Страница 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Страница 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Страница 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Страница 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Страница 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Страница 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Страница 294: ...294 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 17 Digital Voltmeter and Counter...
Страница 300: ...300 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 18 Frequency Response Analysis...
Страница 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Страница 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Страница 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Страница 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...
Страница 416: ...416 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 25 CAN LIN Triggering and Serial Decode...
Страница 426: ...426 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 26 CXPI Triggering and Serial Decode...
Страница 436: ...436 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 27 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 30 Manchester NRZ Triggering and Serial Decode...
Страница 510: ...510 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 32 SENT Triggering and Serial Decode...
Страница 520: ...520 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 33 UART RS232 Triggering and Serial Decode...