30
Keysight N5990A MIPI M-PHY Receiver Test User Guide
2
N5990A M-PHY Station
DSGA
• To use this setup, select
JBERT
from the
Configuration
drop-down and
check the
Separated Low Speed Generator
option.
• This system configuration works similarly to the
M8020A
+
DSGA
configuration. See section
for more information on
the system configuration.
DSGA
• To use this setup, select
DSGA
from the
Configuration
drop-down. The
DSGA
standalone configuration is only supported for transmitter
testing. It sends the training sequence to the DUT and the actual
testing is performed by the scope app.
• For any receiver or transmitter test configuration, having a
DSGA
enables sending hardware reset signals to the DUT. This allows
seamless transition from one test mode to the next without user
intervention.
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......