280
Keysight N5990A MIPI M-PHY Receiver Test User Guide
8
Sys Burst Tests
Figure 161
Example HTML Viewer for Test 2.3.4-SYS-RX Differential Termination En-
able Time (T
TERM-ON-SYS-RX
) Test
Table 51
Test 2.3.4-SYS-RX Differential Termination Enable Time (T
TERM-ON-SYS-RX
) Test data
table
Parameter Name
Parameter Description
Result
"Pass"/"Fail" - If the value is "Fail", the termination enable time is not within the spec limit.
Termination Time Min
It is the shortest/minimum termination enable time in "ns".
Termination UI Min
It is the shortest/minimum termination enable time in "UI".
Min Spec
It is the minimum value of the termination enable time, where the DUT must work according to the
specification.
Termination Time Max
It is the maximum termination enable time in "ns".
Termination UI Max
It is the maximum termination enable time in "UI".
Max Spec
It is the maximum value of the termination enable time, where the DUT must work according to the
specification.
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......