Keysight N5990A MIPI M-PHY Receiver Test User Guide
79
Calibrations and Test Procedures
3
Prepare Length Capability
It is a DUT attribute that determines the minimum PREPARE time supported. It has a permitted range from 0 to
15.
Pulse Distance
It is the distance between DIFN pulses. If only one pulse is applied, it gives the distance between the following
DIFP (Positive Differential Voltage) state next to the DIFN state.
Pulse Width
It is the width of the DIFN pulse. It's default value is 20 ns.
PWM Data Rate
It is the data rate for the PWM mode.
PWM Receive Ratio
It is the PWM Receive Ration provided for the test.
Ref Clock Adjustment
It can be adjusted as “NominalRefClock” and “NominalDataRate”.
Ref Clock Frequency
It is the frequency value of the reference clock.
Reference Clock Skew Step Size
It is the step size value of the reference clock skew during the Jitter Tolerance test.
Reference Clock Skew Step
Duration
It is the step duration value of the reference clock skew during the Jitter Tolerance test.
Re-Init Sequence after Reset DUT
If the value is set to True, the test sequence starts again from the beginning every time the DUT is reset.
RJ (p-p) Amplitude
It is the amplitude value of the RJ added to the test signal during calibration.
Scope RJrms
During RJ calibrations, the oscilloscope adds an additional amount of jitter. This jitter can be calculated by
the scope using EZJit Wizard. If that calculated value is set to the “Scope RJrms” parameter, it will be
subtracted from the RJ measurement.
Semi Automated Test
This property is used for HS, PWM and SYS Termination tests. If false, the software automatically measures the
termination time. If true, the automation finds the spot where the burst starts but lets you place the markers
manually. Then, with the position of the markers, the termination time is calculated. This is useful in cases
where the software is unable to calculate this time automatically because of, for example, reflections in the
signal that impair the algorithm.
Sequence
It is the sequence file used during the test to the DUT.
Set Offset
It is the signal offset set for the data channel.
Set Single Ended Amplitude
It is the single ended amplitude value set for the data channel.
Short Term Deterministic Jitter
It is the short term deterministic jitter amplitude value.
Short Term Random Jitter / STRJ
Amplitude (p-p)
It is the amount of short term random jitter amplitude added to the signal.
Short Term Total Jitter
It is the short term total jitter amplitude value.
Show Real Time Eye
It can be selected as “True”/”False” to switch between real time eye and single waveform. The measurement
time increases by viewing the real time eye.
Sleep Length
It is the provided sleep length value (it is the power saving state between LS Bursts).
Parameter Name
Parameter Description
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......