Keysight N5990A MIPI M-PHY Receiver Test User Guide
261
PWM Tests
7
Procedure
The data generator sends Burst-mode CJTPAT signaling with nominal
amplitude settings while the DUT is configured for Terminated-mode or
Non-Terminated-mode depending on the test case.
The TOL
PWM-G1-LR-RX
is the ratio of the PWM Receive Bit Duration
(T
PWM-RX
) and the average of N PWM receive bit durations during
LINE-READ in PWM-G1. The TOL
PWM-G1-LR-RX
is tested for the
conformance limit values (0.89/1.01) and for the given Additional Steps. At
each step, the BER is measured and compared with the Target BER to
consider the pass/fail result.
This test is performed at each selected lane and PWM gear.
Connection Diagram
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.1- HS-RX Differential Input Voltage Amplitude Test
test.
Parameters
The parameters used in
Expert Mode
for this test, are as follows:
• CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
)
• CDR Peaking (if CDR is selected in the Special Parameter dialog, see
• CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
)
• CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
• Target BER
• IBerReader Init Mode
• Test Sequence
• Re-Init sequence after Reset DUT
• Lane Under Test Termination Model
• Min Tested
• Max Tested
• Additional Steps
• Sequence
These parameters are listed in alphabetical order in
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......