Max Tested HFSJ Amplitude
It is the maximum value of the HFSJ to be tested.
Max Tested LFSJ Amplitude
It is the maximum value of the LFSJ to be tested.
Max Tested Value
It is the maximum skew value to be tested.
Max. Time Span Ch1/2
It is the time span set to the oscilloscope channels.
Minimum Accumulated Differential
Voltage
If the measured accumulated differential voltage is smaller than the given differential voltage, the calibration
stops.
Frequency Offset
It is the frequency deviation relative to the nominal bit rate.
Min Interference Amplitude
It is the minimum value of the interference amplitude to be tested.
Min Interference Value
It is the lower limit for the tested RF Interference. If the minimum and maximum interference values are equal,
only one interference level is tested.
Min Jitter Value
It is the minimum jitter value to be calibrated.
Min Spec
It is the minimum value of the parameter that the DUT must tolerate according to the specification.
Min Tested Value
It is the minimum skew value to be tested.
Min User-Defined Interference
Amplitude
It is a user-defined value for the Common Mode Interference test. It determines the minimum Interference
value that must meet the target BER to pass the test.
Min User-Defined SJ Amplitude
It is a user-defined value for the Jitter Sensitivity test. It determines the minimum SJ amplitude value that
must meet the target BER to pass the test.
Number of Histogram Hits
It is the number of histogram hits using the oscilloscope.
Number of Pulses
The default value is 1, that is, only one DIFN (negative differential voltage) pulse is generated.
Offset Range
It is the range of the offset values to be tested. It is defined with four values separated by semicolons: <Start
Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
Oscilloscope Bandwidth
It allows modification of the oscilloscope bandwidth in calibrations. As it impacts the measured jitter, the
default value is recommended.
Pattern Padding Mode
It can be selected as “Multiply Pattern” and “Fill Pattern”.
Perform Jitter Limit Test
This property is set to to true, in the Jitter Tolerance Test (2.1.7), to search for the maximum amount of jitter
that the DUT can tolerate. The SJ component is increased until the BER test fails.
If is set to false (by default), the total jitter is set to the worst case scenario defined in the CTS.
PJ Amplitude
It is the value of the PJ (Period Jitter) amplitude.
PJ Frequency
It is the frequency value of PJ.
(HS/LS) Prepare Length
It defines the PREPARE sub-state. It specifies the number of DIFP (Differential Positive Pulse) states before the
HS/LS-burst starts. It must be specified in SI (1SI = 10UI).
Parameter Name
Parameter Description
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......