Keysight N5990A MIPI M-PHY Receiver Test User Guide
193
HS Tests
5
Procedure
This test is similar to the previous Test 2.1.8 but in this case, the data
generator sends Continuous-Mode CJTPAT signaling instead of
Burst-Mode, and the DUT is configured to receive the Continuous-Mode
CJTPAT in Terminated mode. For more details refer to the Procedure
section of the Test 2.1.8-HS-RX Frequency Offset Tolerance (fOFFSET-RX)
during HS Burst).
Connection Diagram
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.1- HS-RX Differential Input Voltage Amplitude Test.
Parameters
The parameters used in
Expert Mode
for this test, are as follows:
• CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
)
• CDR Peaking (if CDR is selected in the Special Parameter dialog, see
• CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
)
• CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
• Target BER
• BerReader Init Mode
• Test Sequence
• Lane Under Test Termination Model
• Level Pairs
• Tested Offset
• Tested Range
These parameters are listed in alphabetical order in
Dependencies
The calibrations required for this test are as follows:
• Reference Clock Calibration
• Level Calibration Terminated
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......