Keysight N5990A MIPI M-PHY Receiver Test User Guide
199
HS Tests
5
Figure 121
Example HTML Viewer for Test 2.1.9-HS-RX PREPARE Length Verification
Table 31
Test 2.1.9-HS-RX PREPARE Length Verification data table
Parameter Name
Parameter Description
Result
"Pass"/"Fail" - If the value is "Fail", the DUT does not support the tested prepare length.
Tested HS_Prepare[ ]
It is the value of the tested HS-Prepare length.
Min Passed
It is the shortest prepare length value supported by the DUT.
Min Tested
It is the shortest prepare length that is tested.
Max Passed
It is the maximum prepare length value supported by the DUT.
Max Tested
It is the maximum prepare length value that is tested.
Tested Amplitude
It is the differential voltage amplitude value to be tested.
Tested Offset
It is the common mode voltage value that is tested.