Keysight N5990A MIPI M-PHY Receiver Test User Guide
3
Calibrations and Test Procedures
Group Parameters
Group parameters are used for several related calibration or test
procedures. They are shown on the right side of the N5990A user interface
when the selected entry of the procedure tree on the left is a group instead
of an individual procedure.
The MIPI M-PHY Receiver Test Software has some group parameters (in
addition to Repetitions on the top-level entry of the procedure tree (see
). These will be common for all N5990A procedures.
Figure 37
MPHY Group Parameters
describes the group parameters.
Table 5
Group Parameters
Parameter Name
Parameter Description
0 V Amplitude at Init
The value is True/False. If is set to true, the amplitude of the generator will be set to 0 Volts during the
Init phase of each test.
Reduced Amplitude at Init
The voltage value set during the Init phase of each test.
Re-Init Sequence
The value is True/False. The default value is false. If the value is set to True, the test sequence starts
again from the beginning in each test step to ensure that the DUT is in a defined state. This results in an
increased test duration
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......