104
Keysight N5990A MIPI M-PHY Receiver Test User Guide
4
Calibrations
Embedded Fixture ISI (w/Switch) Calibration
This calibration is available for the M8020A configurations. It is required for
HS Gear3 and Gear 4 when the M8020A ISI is selected as Automatic (see
Figure 59
Embedded Fixture ISI Calibration
Purpose
The receiver test setup must include an ISI Compliance Channel, capable
of introducing a channel loss as defined in the M-PHY specification. This
procedure calibrates the ISI when it is generated with the embedded
M8020A ISI fixture and is selected as Automatic (see
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......