52
Keysight N5990A MIPI M-PHY Receiver Test User Guide
2
N5990A M-PHY Station
Custom BER Reader
The usage of a Custom BER Reader
enables fully automated testing for all
transmission modes (HS and LS) without the need for a BER ED or Unipro
BER reader support. This method requires the implementation of a class
supporting the IBerReader interface by you, providing access to the DUT's
pass/fail information. For example, by reading the DUT's internal error
counter registers. The definition of the IBerReader interface is available
from Keysight.
Address
It is the BER reader's address. It may be an IP address or a file directory.
Get Errors from BERT
This option is enabled if
Custom BER Reader
is chosen. It allows to read
the errors from the BERT Analyser like in Loopback, but controls the rest of
the test execution.
Settings
This option is visible when
Unipro BER Reader
is chosen. Click
Settings
to
open the
Analyzer Settings
dialog as shown in
.
Figure 23
Analyzer Settings
In
Unipro Test Mode
, either the
M8020A
or the
DSGA
can be used as
Analyzer
.
In case of
DSGA
, it is necessary to specify in the
PWM Data Rate
property,
a rough estimate of the data rate the DUT is responding with. The
Compare Mode
can be selected as
Differential
or
Single-Ended
. The
Single-Ended
mode allows you to connect the Tx- to the
DSGA
and Tx+ to
the scope and verify that the DUT is working, by using the scope decoder.
Calibrations Files Directory
Browse
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......