130
Keysight N5990A MIPI M-PHY Receiver Test User Guide
4
Calibrations
When the DSGA is used, two calibrations are required at each data rate:
one for Data0 and another one for the other channels. The calibration for
Data0 includes the switch. For the other channels, Data1 is calibrated
without the switch.
Procedure
First, the Sinusoidal Jitter is adjusted until the measured eye width is
within specified limits. At each step, it starts with the default value and is
then recalculated based on the difference with the target value. Once the
jitter components are set, a sweep over the differential voltage is
performed until the measured eye height is below the specified. Up to
gen3, the eye width and height values are obtained by using the histogram
measurements of the DSO. For gen 4, the jitter measurement are
performed with the SigTest software.
This procedure is repeated for different frequencies: f
SJ0_RX
, f
SJ1_RX
,
f
SJ2_RX
, f
SJ3_RX
, f
SJ4_RX
defined in the specifications
Connection Diagram
Refer to the connection diagrams shown in the Connection Diagram
section of the ISI (w/Switch) Calibration test for M8020A, and N4903B
configurations.
Parameters
The parameters used in
Expert Mode
for this calibration, are as follows:
• Lane Under Test Termination Mode
• Oscilloscope Bandwidth
• Jitter Frequencies
• Eye Width Target
• Eye Width Max Variation
• Eye Height Target
NOTE
If the calibration fails and shows messages such as "The Eye -Height did
not reach a value above the target value of xx mV!" and "Check your
system connections and if necessary increase the
Maximum Calibration
Voltage
in Calibration properties, try to increase the
Maximum
Calibration Voltage
amplitude value and run the calibration again.
Repeat this step until it passes successfully.
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......