Keysight N5990A MIPI M-PHY Receiver Test User Guide
65
Calibrations and Test Procedures
3
Example of MIPI M-PHY Calibration / Test Procedure
All calibration and test procedures are included in the respective groups in
a way similar to how they are organized in the CTS. For most procedures,
some specific parameters can be set in
Expert Mode
by the user. In
Level Calibration Into Open
is highlighted as an example.
The respective parameters are shown on the right side of the N5990A User
Interface. This is achieved by clicking on the calibration/test name. To start
one or more procedures, do the following:
1 Select the check box for the corresponding procedure. The
Start
button is enabled and colored in green.
2 Click
Start
to run the procedure.
3 Once all the procedures are run, click
Save
to store the N5990A
configuration as a single ".vfp" file. You can also click
Load
to recall
an N5990A configuration. In this way, you can run the DUT again
without configuring it.
Figure 30
Example of MIPI M-PHY Calibration or Test Procedure